Ventilation
Trilogy 100 Ventilator
This product is no longer available

Trilogy 100

Ventilator

This product is no longer available

Find similar products

Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute- care use to provide invasive and non-invasive ventilator support for adult and pediatric patients. A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology.

Specifications

Alarms
Alarms
Apnea rate
  • Off; 10 to 60 s
Circuit disconnect
  • Off; 5 to 60 s
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 40 to 2000 ml
High minute ventilation
  • Off; 1.0 to 99 l/min
Low minute ventilations
  • Off; 0.2 to 99 l/min
High respiratory rate
  • Off; 4 to 80 BPM
Low respiratory rate
  • Off; 4 to 80 BPM
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 BPM
Minute ventilation
  • 0 to 99 l/min
Vte/Vti
  • 0 to 2000 ml
Leak rate
  • 0 to 200 l/min
Peak inspiratory flow
  • 0 to 200 l/min
Peak inspiratory pressure
  • 0 to 99 cm H₂O
Mean airway pressure
  • 0 to 99 cm H₂O
% Patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Dimensions and weight
Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
  • 5 kg (11 lb)
Control accuracy
Control accuracy
Circuit types
  • Active with PAP; Passive
Tidal volume
  • 50 to 2000 ml
Breath rate
  • 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Ventilation types
  • Pressure and Volume
Inspiratory time
  • 0.3 to 5.0 s
Volume modes
  • AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
  • CPAP; S; S/T; T; PC; PC-SIMV (w/PS); PC-MPV; AVAPS-AE
Rise time
  • 1 to 6
Pressure
  • 4 to 50 cm H₂O
IPAP
  • 4 to 50 cm H₂O
Ramp start pressure
  • 0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP
Ramp length
  • Off; 5 to 45 minute(s)
PEEP/EPAP for active circuits
  • 0 to 25 cm H₂O
PEEP/EPAP for passive circuits
  • 4 to 25 cm H₂O
Flex
  • Off; 1 to 3
CPAP for passive circuits
  • 4 to 20 cm H₂O
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 BPM
Pressure support differential
  • 0 to 40 cm H₂O
Alarms
Alarms
Apnea rate
  • Off; 10 to 60 s
Circuit disconnect
  • Off; 5 to 60 s
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 BPM
Minute ventilation
  • 0 to 99 l/min
See all specifications
Alarms
Alarms
Apnea rate
  • Off; 10 to 60 s
Circuit disconnect
  • Off; 5 to 60 s
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 40 to 2000 ml
High minute ventilation
  • Off; 1.0 to 99 l/min
Low minute ventilations
  • Off; 0.2 to 99 l/min
High respiratory rate
  • Off; 4 to 80 BPM
Low respiratory rate
  • Off; 4 to 80 BPM
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 BPM
Minute ventilation
  • 0 to 99 l/min
Vte/Vti
  • 0 to 2000 ml
Leak rate
  • 0 to 200 l/min
Peak inspiratory flow
  • 0 to 200 l/min
Peak inspiratory pressure
  • 0 to 99 cm H₂O
Mean airway pressure
  • 0 to 99 cm H₂O
% Patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Dimensions and weight
Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
  • 5 kg (11 lb)
Control accuracy
Control accuracy
Circuit types
  • Active with PAP; Passive
Tidal volume
  • 50 to 2000 ml
Breath rate
  • 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Ventilation types
  • Pressure and Volume
Inspiratory time
  • 0.3 to 5.0 s
Volume modes
  • AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
  • CPAP; S; S/T; T; PC; PC-SIMV (w/PS); PC-MPV; AVAPS-AE
Rise time
  • 1 to 6
Pressure
  • 4 to 50 cm H₂O
IPAP
  • 4 to 50 cm H₂O
Ramp start pressure
  • 0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP
Ramp length
  • Off; 5 to 45 minute(s)
PEEP/EPAP for active circuits
  • 0 to 25 cm H₂O
PEEP/EPAP for passive circuits
  • 4 to 25 cm H₂O
Flex
  • Off; 1 to 3
CPAP for passive circuits
  • 4 to 20 cm H₂O
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 BPM
Pressure support differential
  • 0 to 40 cm H₂O

You are about to visit a Philips global content page

Continue

Our site can best be viewed with the latest version of Microsoft Edge, Google Chrome or Firefox.